HPLC Method for Determination of Vecuronium on Primesep SB Column

HPLC-Applikation • Datum: 2022/09/13
HPLC Method for Determination of Vecuronium on Primesep SB Column Chromatogramm

High Performance Liquid Chromatography (HPLC) Method for Analysis of Vecuronium bromide

Vecuronium bromide , also known as Norcuron , is a muscle relaxant with the chemical formula C34H57BrN2O4 . It is primarily used during surgery and executions by lethal injection. Reversal of vecuronium’s effects can be achieved through administration sugammadex.

HPLC methods for Vecuronium bromide are of great interest to pharmaceutical research and industry. Vecuronium bromide can be separated and retained on a mixed-mode stationary phase Primesep SB column with embedded strong basic ion-pairing groups, using an isocratic analytical method with a simple mobile phase of water, Acetonitrile (MeCN), and an Ammonium Formate (AmFm) ionic modifier. This analysis method can be UV detected at 260 nm with high resolution and peak symmetry and is compatible with Mass Spectrometry (MS), ELSD, and CAD.

Methodenparameter

ColumnPrimesep SB, 2.1 x 100 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN – 10%
BufferAmmonium Formate pH 3.0 – 50 mM, 100 mM
Flow Rate0.2 mL/min
DetectionELSD, 50C