HPLC Determination of Mycophenolate Mofetil on Primesep SB Column

HPLC-Applikation • Datum: 2019/05/01
HPLC Determination of Mycophenolate Mofetil on Primesep SB Column Chromatogramm

High Performance Liquid Chromatography (HPLC) Method for Analysis of Mycophenolate mofetil .

Mycophenolate mofetil alongside other medicines is used as an immunosuppressant to prevent organ rejection after a transplant. THe compound can be retained in HPLC using Primesep SB mixed-mode column that has both hydrophobic and anion-exchange properties. The analytical method’s mobile phase consists of a gradient of acetonitrile (ACN) in water with phosphoric acid (H3PO4) buffer and using UV detection at 250 nm.

Methodenparameter

ColumnPrimesep SB, 4.6 x 150 mm, 5 µm, 100 Å, dual ended
Mobile PhaseGradient MeCN – 10-20%, 10 min
BufferH3PO4 – 0.2%
Flow Rate1.0 mL/min
DetectionUV, 250 nm