HPLC Determination of Erioglaucine on Newcrom BH Column

HPLC-Applikation • Datum: 2020/07/02
HPLC Determination  of Erioglaucine on Newcrom BH Column Chromatogramm

High Performance Liquid Chromatography (HPLC) Method for Analysis of Brilliant Blue FCF (Blue 1) .

Erioglaucine, more commonly known as Brilliant Blue FCF, is a synthetic coloring dye used in many industries as a colorant for foods, medications, cosmetics, etc. It is a multi-charged molecule and generally has poor peak shape on reverse-phase HPLC columns. By using a Newcrom BH mixed-mode column with both ion-exchange and hydrophobic properties, erioglaucine can be retained and exhibits great peak shape by using a mobile phase of methanol (MeOH) and water with sulfuric acid as buffer. UV detection at 625 nm.

You can find detailed UV spectra of Brilliant Blue and information about its various lambda maxima by visiting the following link.

Methodenparameter

ColumnNewcrom BH, 4.6 x 50 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeOH/H2O – 50/50%
BufferH2SO4 – 1%
Flow Rate1.0 mL/min
DetectionVis 625 nm