HPLC Determination of Allyl Bromide on Newcrom R1 Column

HPLC-Applikation • Datum: 2019/11/07
HPLC Determination of Allyl Bromide on Newcrom R1 Column Chromatogramm

High Performance Liquid Chromatography (HPLC) Method for Analysis of Allyl bromide .

Allyl Bromide is an organic halide with C3H5Brchemical formula. It is an alkylating agent typically found in synthesis of polymers, pharmaceuticals, perfumes, and more. Often, it is used as starting material in organic synthesis. Physically, it is a colorless liquid that occasionally presents more yellow. It is toxic to people by skin absorption, irritable to eyes and respiratory system.

Allyl bromide can be retained and analyzed using the Newcrom R1 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a triethanolamine buffer. Detection is performed using UV.

Methodenparameter

ColumnNewcrom R1, 4.6 x 150 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN/H2O – 60/40%
Flow Rate1.0 mL/min
DetectionUV 210 nm