Alltesta HPLC Method for Analysis of Linear PEI + (Cu) on PEI Column

HPLC-Applikation • Datum: 2024/05/28
Alltesta HPLC Method for Analysis of Linear PEI + (Cu) on PEI Column Chromatogramm

Separation type: Liquid Chromatography Reversed-phase

Polyethyleneimine (PEI) is a polymer used in a wide range of applications. It is difficult to analyze in HPLC due to several factors. It’s a mixture of compounds with different lengths and therefore different number of charges. It also lacks a UV chromophore. New PEI column was designed specifically for the analysis of polyethyleneimine by ion-exclusion and size-exclusion mechanisms with copper complex to allow for UV detection. The method uses a mobile phase of acetonitrile (ACN) and water with ammonium formate buffer (AmFm) and UV detection at 275 nm

Methodenparameter

ColumnPEI, 4.6 x 250 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN/H2O – 40/60%,
BufferAmmonium formate pH 3.0 – 20 mM
Flow Rate0.5 mL/min
DetectionUV 275 nm