HPLC Determination of Oxalic Acid on Newcrom B Column

HPLC-Applikation • Datum: 2019/11/05
HPLC Determination of Oxalic Acid on Newcrom B Column Chromatogramm

High Performance Liquid Chromatography (HPLC) Method for Analysis of Oxalic Acid

Oxalic Acid . also known as ethanedioic acid is an organic acid. It occurs naturally in plants, but is also often found in industrial uses as a bleaching agent, rust remover, and cleaning agent. Industrial uses of it also extend into households, as it is easily accessible. Despite that, in large amounts, it can be toxic.

Oxalic Acid can be retained and analyzed using the Newcrom B stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a formic acid buffer. Detection is performed using CAD.

Methodenparameter

ColumnNewcrom B, 4.6 x 150 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN/H2O – 20/80%
BufferFormic Acid – 2.0%
Flow Rate1.0 mL/min
DetectionCAD (Corona)