HPLC Separation of DOPAC, Homovanillic Acid and Dopamine on the Primesep 100 Column

HPLC-Applikation • Datum: 2012/07/10
HPLC Separation of DOPAC, Homovanillic Acid and Dopamine on the Primesep 100 Column Chromatogramm

Application Notes: Neurotransmitters DOPAC, homovanillic acid and dopamine were separated by mixed-mode chromatography on Primesep 100 and Primesep 200 HPLC columns. The method can be used for quantification of neurotransmitters with LC/MS compatible conditions. The compounds are retained by combination of reversed-phase, ion-exchange or ion-exclusion mechanisms. The retention time and selectivity of separation can be adjusted by variation of amount of acetonitrile, buffer pH and buffer concentration.

Application Columns: Primesep 100

Application compounds: Dopac, Homovanillic Acid, Dopamine

Detection technique: UV, LC/MS, ELSD/CAD

Methodenparameter

ColumnPrimesep 100, 4.6×50 mm, 5 µm, 100 Å
Mobile PhaseMeCN/H2O – 5/95%
BufferAmFm pH 3.0- 30 mM
Flow Rate1.0 mL/min
DetectionUV, 270 nm