HPLC Method for Separation of Sodium, Potassium, Lithium, and Ammonium on Primesep 100 Column

HPLC-Applikation • Datum: 2024/08/07
HPLC Method for Separation of Sodium, Potassium, Lithium, and Ammonium on Primesep 100  Column Chromatogramm

Separation type: Liquid Chromatography Mixed-mode

Inorganic cations and ammonia are usually analyzed by ion-exchange chromatography with a conductivity detector. Mixed-mode chromatography achieves the same goal with an ELSD detector. In addition to retaining ions, mixed-mode column can retain hydrophobic molecules also, making it possible to analyze a wide variety of ionic and non-ionic compounds in one run. In this method sodium, potassium, and lithium ions were separated along with ammonium ions on a Primesep 100 column. Detection technique is ELSD.

Methodenparameter

ColumnPrimesep 100, 4.6 x 250 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN/H2O – 80/20%
BufferTFA – 0.1%
Flow Rate1.0 mL/min
DetectionELSD, the nebulizer and evaporator temperatures 50°C, with a gas flow rate of 1.6 Standard Liters per Minute (SLM)