HPLC Method for Analysis of Chromium(III) on BIST B+ Column

HPLC-Applikation • Datum: 2023/02/06
HPLC Method for Analysis of Chromium(III) on BIST B+ Column Chromatogramm

Separation type: Bridge Ion Separation Technology, or BIST™ by SIELC Technologies

Chromium(III) is a nutrient essential for human metabolism of insulin, sugar, and lipids. Using SIELC’s newly introduced BIST™ method, Cr(III) can be retained on a positively-charged anion-exchange BIST™ B+ column. There are two keys to this retention method: 1) a multi-charged, negative buffer, such as Sulfuric acid (H2SO4), which acts as a bridge, linking the positively charged trivalent Chromium ion to the positively charged column surface and 2) a mobile phase consisting mostly of organic solvent (such as MeCN) to minimize the formation of a solvation layer around the charged analytes. Using this new and unique analysis method, Chromium(III) can be separated, retained, and UV detected at 200 nm.

Methodenparameter

ColumnBIST B+, 4.6 x 50 mm, 5 µm, 100 Å, dual ended
Mobile PhaseGradient MeCN – 60-20%, 10 min
BufferH2SO4 – 0.2%
Flow Rate1.0 mL/min
DetectionUV 200 nm
Limit of Detection0.4 ppm