HPLC Method for Analysis of 4-Nitrobenzoyl Chloride on Primesep 200 Column

HPLC-Applikation • Datum: 2026/02/20
HPLC Method for Analysis of 4-Nitrobenzoyl Chloride on Primesep 200 Column Chromatogramm

High Performance Liquid Chromatography (HPLC) Method for Analysis of Epibatidine

4-nitrobenzoyl chloride is an acyl halide with the chemical formula C7H4ClNO3 . It is primarily used in preparation of polysubstituted furanonaphthoquinoines and as an intermediate in active pharmaceutical ingredients and dyes. You can find detailed UV spectra of 4-Nitrobenzoyl chloride and information about its various lambda maxima by visiting the following link.

4-Nitrobenzoyl chloride can be retained and analyzed using the Primesep 200 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a sulfuric acid buffer. Detection is performed using UV.

Methodenparameter

ColumnPrimesep 200, 4.6 x 150 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN/H2O – 40/60%
BufferH2SO4 – 0.1%
Flow Rate1.0 mL/min
DetectionUV 200 nm