ELSD HPLC Method for Analysis of Captopril on Newcrom B Column

HPLC-Applikation • Datum: 2025/01/22
ELSD HPLC Method for Analysis of Captopril on Newcrom B Column Chromatogramm

Captopril is a widely used angiotensin-converting enzyme (ACE) inhibitor, effective in managing high blood pressure (hypertension), heart failure, and diabetic kidney complications. By relaxing blood vessels and enhancing blood flow, it reduces strain on the heart and promotes cardiovascular health.

Captopril can be retained, and analyzed using the Newcrom B mixed-mode stationary phase column. The analysis employs an isocratic method with a simple mobile phase consisting of water, acetonitrile (MeCN), and ammonium formate as a buffer. Detection is performed using an Evaporative Light Scattering Detector (ELSD).

*LOD was determined for this combination of instrument, method, and analyte, and it can vary from one laboratory to another even when the same general type of analysis is being performed.

Methodenparameter

ColumnNewcrom B, 4.6 x 150 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN10%
BufferAmmonium Formate pH 3.0 – 20 mM
Flow Rate1 mL/min
DetectionELSD, the nebulizer and evaporator temperatures 40°C, with a gas flow rate of 1.6 Standard Liters per Minute (SLM)