HPLC Method for Analysis of Alachlor on Primesep 100 Column

https://sielc.com/hplc-method-for-alachlor
⬇ PDF Download

Chromatogram

Description

· Separation type: Liquid Chromatography Mixed-mode SIELC Technologies · HPLC Method for Analysis of Alachlor on Primesep 100 Column

Alachlor is a chloroacetanilide herbicide used to control annual grasses and broadleaf weeds primarily in corn and soybean crops. It is known for its pre-emergent activity, meaning it is applied to the soil to prevent weed germination and growth.

Alachlor be retained and analyzed using a Primesep 100 mixed-mode stationary phase column. The analysis employs an isocratic method with a simple mobile phase comprising water, acetonitrile (MeCN), and sulfuric acid as a buffer. This method allows for detection using UV 200 nm

Method Parameters

Mobile PhaseMeCN – 55%
BufferH2SO4 -0.1%
Flow Rate1.0 ml/min
DetectionUV 200 nm
Samples1.0 mg/ml in MeCN – 100%
Injection volume1 µl
LOD*8 ppb (200 nm)
Class of Compoundschloroacetanilide herbicides
Analyzing CompoundsAlachlor

HPLC Column Used

Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended