Application Notes: Guaifenesin is common, over the counter expectorant. Guaifenesin contain no less than 98 percent and not more than 102 percent of the labeled amount of guaifenesin calculated on a dried basis, according to the USP methods. the The USP HPLC method for the analysis of guaifenesin was developed on our Legacy L1 column according to the US Pharmacopeia methodology. L1 classification is assigned to reversed-phase HPLC column containing C18 ligand. Support for the material is spherical silica gel with particles size 3-10 µm and pore size of 100-120 Å.
Application Columns: Legacy L1 C18HPLCcolumn
Application compounds: Guaifenesin, benzoic acid
Mobile phase: MeOH/H2O/AcOH40:60:1.5
Detection technique: UV
Reference: USP35- NF30
| Column | Legacy L1, 4.6×150 mm, 5 µm, 100 Å |
|---|---|
| Mobile Phase | MeOH/H2O/AcOH40/60/1.5 |
| Buffer | NaH2PO4 |
| Flow Rate | 1.0 mL/min |
| Detection | UV, 270 nm |