USP Method for the Analysis of Amoxicillin using the Legacy L1 Column

HPLC-Applikation • Datum: 2012/06/21
USP Method for the Analysis of Amoxicillin using the Legacy L1 Column Chromatogramm

Application Notes: Amoxicillin is one of the most commonly prescribed antibiotics. It is often used for treating strep throat. According to USP methods, amoxicillin should not contain less than 900ug and no more than 1050ug of amoxicillin per mg. The USP HPLC method for the separation of amitriptyline was developed on Legacy L1 column according to the US Pharmacopeia methodology. L1 classification is assigned to reversed-phase HPLC column containing C18 ligand. Support for the material is spherical silica gel with particles size 3-10 µm and pore size of 100-120 Å. Resolution between critical pairs corresponds to rules and specifications of UPS.

Application Columns: Legacy L1 C18HPLCcolumn

Application compounds: Amoxicillin

Mobile phase: MeCN/50 mM dibasic sodium phosphate

Detection technique: UV

SIELC's family of Legacy columns is based on the United States Pharmacopeia's (USP) published chromatographic methods and procedures. Numerous brands have columns used in USP reference standards and methods. USP has created various designations to group together columns with similar types of packing and properties in the solid phase. SIELC's Legacy columns adhere to these strict requirements and properties, allowing you to easily replace older columns that are no longer available without needing to significantly modify your method or SOPs.

Methodenparameter

ColumnLegacy L1, 4.6×150 mm, 5 µm, 100 Å
Mobile PhaseMeCN/50 mM NaH2PO4pH 5.0 (4/96)
BufferNaH2PO4
Flow Rate1.0 mL/min
DetectionUV, 230 nm