Separation of Maleic and Succinic Acid on Primesep B2 column

HPLC-Applikation • Datum: 2011/07/08
Separation of Maleic and Succinic Acid on Primesep B2 column Chromatogramm

Organic and inorganic acids can be retained and separated on mixed-mode columns based on weak reversed-phase and weak/medium anion-exchange mechanisms. Amount of ACN, buffer concentration and buffer pH will affect retention time of organic and inorganic acids. Acids can be monitored by low UV, ELSD or LC/MS. Presence of ions is required to facilitate ion-exchange mechanism. Method can be used as a general approach for analysis of acidic hydrophilic and acidic hydrophobic compounds. Carboxylic acids along with inorganic acid can be retained and separated without ion-pairing reagent.

Methodenparameter

ColumnPrimesep B2, 4.6×50 mm, 5 µm, 100 Å
Mobile PhaseMeCN/H2O – 10/90%
BufferAmFm pH 3.0
Flow Rate1.0 mL/min
DetectionELSD, UV 230 nm