HPLC Separation of Acrylamide and Melamine on Primesep 200 Column

HPLC-Applikation • Datum: 2020/09/04
HPLC Separation of Acrylamide and Melamine on Primesep 200 Column Chromatogramm

High Performance Liquid Chromatography (HPLC) Method for Analysis of Acrylamide , Melamine .

Melamine is a nitrogen-based organic compound that was first synthesized by the German chemist Justus von Liebig in 1834. Melamine is a decorative coating that is resistant to water and mechanical damage.

Melamine is used for:

We were able to separate Melamine from Acrylamide in HPLC on Primesep 200 column using a mobile phase of water with Formic Acid buffer. This separation can be monitored with UV, ELSD, CAD detection techniques.

Methodenparameter

ColumnPrimesep 200, 4.6 x 150 mm, 3 µm, 100 Å, dual ended
Mobile PhaseH2O – 99%
BufferFormic Acid – 1%
Flow Rate1.0 mL/min
DetectionUV 240 nm