HPLC Separation of 2-Fluorophenylacetic acid, 4-Fluorophenylacetic acid, 3-Fluorophenylacetic acid on Primesep SB Column

HPLC-Applikation • Datum: 2021/09/29
HPLC Separation of 2-Fluorophenylacetic acid, 4-Fluorophenylacetic acid, 3-Fluorophenylacetic acid  on Primesep SB Column Chromatogramm

High Performance Liquid Chromatography (HPLC) Method for Analysis of 2-Fluorophenylacetic acid , 4-Fluorophenylacetic acid, 3-Fluorophenylacetic acid

2-Fluorophenylacetic acid is a popular chiral derivatizing agent used in 19F NMR spectroscopy. 3-Fluorophenylacetic acid is a common precursor for synthesizing various pentaamine and bis-heterocyclic compounds. 4-Fluorophenylacetic acid is a common intermediary compound used to produce fluorinated anesthetics. While these 3 compounds all have the same chemical formula, the position of the Fluoride group varies between them, changing slightly their retention and chromatographic characteristics.

All 3 isomers can be measured at low UV. Using a Primsep SB reverse-phase columns and a mobile phase consisting of acetonitrile (MeCN) and water with either formic acid or acetic acid buffer, these fluorophenylacetic acid isomers can be retained, separated, and UV detected at 210 nm with precise resolution.

Methodenparameter

ColumnPrimesep SB, 4.6×250 mm, 5 µm, 100 Å
Mobile PhaseMeCN/H2O – 40/60%
BufferH3PO4 – 0.1%
Flow Rate1.0 mL/min
DetectionUV 210 nm