HPLC Method for Separation of Biopterin, Rhamnopterin and Neopterin on Primesep 100 Column

HPLC-Applikation • Datum: 2023/06/07
HPLC Method for Separation of Biopterin, Rhamnopterin and Neopterin on Primesep 100 Column Chromatogramm

Separation type: Liquid Chromatography Mixed-mode

These three pterins can be retained, separated and analyzed using a reverse-phase Primesep 100, 4.6 x 150 mm, 5 µm, 100 Å, dual ended column. The mobile phase for this method consists of water, acetonitrile (MeCN), and ammonium formate, which serves as a buffer. This analytical method can be monitored using UV detection at 275 nm, an Evaporative Light Scattering Detector (ELSD), or any other evaporative detection method such as Charged Aerosol Detection (CAD) or Electrospray Ionization Mass Spectrometry (ESI-MS)

LOD was determined for this combination of instrument, method, and analyte, and it can vary from one laboratory to another even when the same general type of analysis is being performed.

Methodenparameter

ColumnPrimesep 100, 4.6 x 150 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN/H2O -5/95%
BufferAmmonium formate pH 3.0 – 20 mM
Flow Rate1.0 mL/min
DetectionUV 275 nm
Limit of Detection20 ppb
Injection Volume5 µl