High Performance Liquid Chromatography (HPLC) Method for Analysis of Difluoroacetic acid , Dichloroacetic acid , Dibromoacetic acid , Acetic Acid , Bromoacetic acid , Chloroacetic acid , Bromochloroacetic acid .

Acetic acid and its various chloro- and bromo- forms are widely used in organic chemistry.  Their similar structure makes the acids difficult to retain and separate on reverse-phase HPLC columns.  By using Newcrom BH mixed-mode column which also has ion-exchange properties, the separation can be achieved with a simple isocratic method and relatively short time with a mobile phase of acetonitrile (ACN), water and sulfuric acid (H2SO4) buffer.  UV detection at 200nm.

Method Parameters

Mobile PhaseMeCN/H2O – 10/90%
BufferH2SO4 – 0.3%
Flow Rate1.0 ml/min
DetectionUV 200nm
Class of CompoundsAcid
Analyzing CompoundsDifluoroacetic acid,Dichloroacetic acid,Dibromoacetic acid,Acetic Acid,Bromoacetic acid,Chloroacetic acid,Bromochloroacetic acid

HPLC Column Used

Newcrom BH, 4.6 x 150 mm, 5 µm, 100 A, dual ended