HPLC Method for Separation of Benzonitrile, Toluene and Benzylamine on Primesep A Column

https://sielc.com/hplc-separation-of-aromatics
⬇ PDF Download

Chromatogram

Description

· Separation type: Liquid Chromatography Mixed-mode

These chemicals are widely used in organic synthesis. Please handle these substances with care, as they can have hazardous effects. Always refer to the safety data sheets for each chemical and adhere to recommended safety procedures.

These three compounds be separate and analyzed on a reverse-phase Primesep A, 4.6 x 150 mm, 5 µm, 100 A, dual ended column with a mobile phase consisting of water, Acetonitrile (MeCN) and Sulfuric acid as a buffer modifier. This analysis method can be UV detected at 210 nm.

LOD was determined for this combination of instrument, method, and analyte, and it can vary from one laboratory to another even when the same general type of analysis is being performed.

Method Parameters

Mobile PhaseMeCN/H2O -60/40%
BufferH2SO4
Flow Rate1.0 ml/min
DetectionUV 210 nm
Peaks Retention Time2.9, 4.0, 10.2 min
Samples concentration(Benzonitrile, Toluene, Benzylamine)0.04, 0.05, 0.07 mg/ml
Injection volume3 µl
Sample diluentMeCN/H2O -60/40%
LOD(Benzonitrile, Toluene, Benzylamine)27, 65, 103 ppb
Class of CompoundsBenzene, Aromatics
Analyzing CompoundsBenzonitrile,Toluene,Benzylamine

HPLC Column Used

Primesep A, 4.6 x 150 mm, 5 µm, 100 A, dual ended