HPLC-MS Method for Analysis of Semaglutide on Primesep SB Column

https://sielc.com/hplc-ms-method-for-analysis-semaglutide-primesep-sb
⬇ PDF Download

Chromatogram

Description

· HPLC Method for Analysis of Semaglutide on Primesep SB Column by SIELC Technologies

Semaglutide is a GLP-1 receptor agonist used for managing type 2 diabetes and obesity. It mimics the effects of the natural hormone GLP-1, promoting insulin secretion, reducing glucagon levels, slowing gastric emptying, and suppressing appetite. These mechanisms help control blood sugar levels and support weight loss.

Semaglutide can be retained and analyzed using the Primesep SB stationary phase column. The analysis utilizes a gradient method with a simple mobile phase consisting of water, acetonitrile (MeCN), and Ammonium formate. Detection was performed using UV absorption at 280 nm. Mass spectrometric detection was carried out in ESI-SIM mode, monitoring the ions [M+3H]³⁺ at m/z 1372 and [M+4H]⁴⁺ at m/z 1029

Method Parameters

Mobile PhaseGradient MeCN 40-80%, 10 min
BufferAmmonium Formate pH 3.0 – 40 mM
Flow Rate0.5 ml/min
DetectionUV 280 nm, ESI-SIM: [M+3H]3⁺ -1372, [M+4H]4⁺ -1029
LOD3 ppb
Class of CompoundsGLP-1 Receptor Agonist
Analyzing CompoundsSemaglutide

HPLC Column Used

Primesep SB, 3.2 x 100 mm, 3 µm, 100 A, dual ended