HPLC Method for Analysis of Acetazolamide in Pharmaceutical Formulation on Primesep 100 Column

https://sielc.com/hplc-ms-analysis-of-acetazolamide
⬇ PDF Download

Chromatogram

Description

· HPLC Method for Analysis of Acetazolamide in on Primesep 100 Column

Acetazolamide is a chiral aromatic amine with the molecular formula C 4 H 6 N 4 O 3 S 2 . It is used primarily to treat glaucoma, epilepsy, acute mountain sickness, heart failure, and more. It works as a carbonic anhydrase inhibitor, which causes the accumulation of carbonic acid, which is found in a wide variety of tissue. It is on the World Health Organization’s List of Essential Medicines. You can find detailed UV spectra of [ compounds ] and information about its various lambda maxima by visiting the following link .

Acetazolamide can be retained and analyzed using the Primesep 100 stationary phase column. The analysis utilizes an isocratic method with a mobile phase consisting of acetonitrile, water, and Ammonium Formate. Detection is performed using UV.

Method Parameters

Mobile PhaseMeCN/H2O -40/60%
BufferH2SO4– 0.2%
Flow Rate1.0 ml/min
DetectionUV 266 nm
LOD*20 ppb
Class of CompoundsDrug
Analyzing CompoundsAcetazolamide

HPLC Column Used

Primesep 100, 4.6 x 100 mm, 5 µm, 100 A, dual ended