HPLC Method of Ferrozine on Newcrom BH Column

https://sielc.com/hplc-method-of-ferrozine-on-newcrom-bh-column
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Ferrozine

Ferrozine is a common reagent and indicator used for analyzing and detecting iron. Multi-charged molecules, such as Ferrozine, generally have a tendency to exhibit poor peak shape on reverse-phase HPLC columns, where they show significant tailing.

This compound can be retained with great peak shape on a Newcrom BH mixed-mode column. Ferrozine can be measured at low UV. Using a Newcrom BH mixed-mode column and a mobile phase consisting of methanol (MeOH) and water with perchloric acid (HCLO4) buffer, Ferrozine can be retained and UV detected at 280 nm with precise resolution.

Method Parameters

Mobile PhaseMeOH/H2O – 30/70%
BufferHClO4 – 0.3%
Flow Rate0.2 ml/min
DetectionUV 280  nm
Class of CompoundsIndicator
Analyzing CompoundsFerrozine

HPLC Column Used

Newcrom BH, 2.1 x 100 mm, 5 µm, 100 A, dual ended