HPLC Method for Analysis of Chromium(III) on BIST B+ Column

https://sielc.com/hplc-method-of-chromium
⬇ PDF Download

Chromatogram

Description

· HPLC Method for Analysis of Chromium(III) on BIST B+ by SIELC Technologies

Chromium(III) is a nutrient with the chemical formula Cr 3+ . It is believed to be essential for human metabolism of insulin, sugar, and lipids. Though, the notion is currently contested. It can be found naturally in rocks, soil, plants, as well as various animal products.

Using SIELC’s newly introduced BIST™ method, Chromium(III) can be retained on a positively-charged anion-exchange BIST B+ column. There are two keys to this retention method: 1) a multi-charged, negative buffer, such as Sulfuric acid (H2SO4), which acts as a bridge, linking the positively charged trivalent Chromium ion to the positively charged column surface and 2) a mobile phase consisting mostly of organic solvent (such as MeCN) to minimize the formation of a solvation layer around the charged analytes. Using this new and unique analysis method, Chromium(III) can be separated, retained, and UV detected at 200 nm.

Method Parameters

Mobile PhaseGradient MeCN – 60-20%, 10 min
BufferH2SO4 – 0.2%
Flow Rate1.0 ml/min
DetectionUV 200 nm
Peak Retention Time5.71 min
LOD0.4 ppm
Class of CompoundsIon
Analyzing CompoundsChromium(III)

HPLC Column Used

BIST B+, 4.6 x 50 mm, 5 µm, 100 A, dual ended