HPLC Method for the Analysis of Tirzepatide in Pharmaceutical Formulation Mounjaro on Primesep B Column

https://sielc.com/hplc-method-for-analysis-tirzepatide
⬇ PDF Download

Chromatogram

Description

Tirzepatide is a synthetic peptide that functions as a dual agonist of the GLP-1 and GIP receptors. It is used in the treatment of type 2 diabetes and obesity. By mimicking the actions of both glucagon-like peptide-1 (GLP-1) and gastric inhibitory polypeptide (GIP), tirzepatide enhances insulin secretion, reduces blood sugar levels, promotes satiety, and aids in weight loss.

Tirzepatide can be retained and analyzed using the Primesep B stationary phase column. The analysis utilizes an gradientc method with a simple mobile phase consisting of water, acetonitrile (MeCN), and TFA. Detection is performed using UV at 280 nm.

Method Parameters

Mobile PhaseGradient MeCN – 30-70%, 10 mn
BufferTFA – 0.1%
Flow Rate0.5 ml/min
DetectionUV 280 nm
LOD24 ppb
Class of CompoundsGLP-1 Receptor Agonist
Analyzing CompoundsTirzepatide

HPLC Column Used

Primesep B, 3.2 x 50 mm, 5 µm, 100 A, dual ended