HPLC Method for Analysis of Tetracycline on Primesep 100 Column on Alltesta™

https://sielc.com/hplc-method-for-analysis-of-tetracycline-2
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Tetracycline

Tetracycline is an antibiotic with the molecular formula C 22 H 24 N 2 O 8 . It is a bacteriostatic agent, therefore it prevents bacteria from growing and multiplying through inhibiting protein synthesis within bacteria. It is used to treat a variety of infections.

Tetracycline can be retained and analyzed using the Primesep 100 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a sulfuric acid buffer. Detection is performed using UV.

Method Parameters

Mobile PhaseGradient MeCN – 20-80%
BufferH2SO4 – 0.2%
Flow Rate1.0 ml/min
DetectionUV 275 nm
Class of CompoundsMedication
Analyzing CompoundsTetracycline

HPLC Column Used

Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended