HPLC Method for Analysis of Potassium Sorbate on Newcrom BH Column

https://sielc.com/hplc-method-for-analysis-of-potassium-sorbat
⬇ PDF Download

Chromatogram

Description

High Performance Liquid Chromatography (HPLC) Method for Analysis of Potassium sorbate , Sorbic acid .

Potassium Sorbate is a popular preservative with the chemical formula C 6 H 7 K O 2 . It is used in a wide variety of foods, wines, and personal-care products. It works through inhibiting molds, yeasts, and various other microbes. You can find detailed UV spectra of Potassium Sorbate and information about its various lambda maxima by visiting the following link.

Potassium sorbate , Sorbic acid can be retained and analyzed using the Newcrom BH stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a sulfuric acid buffer. Detection is performed using UV.

Method Parameters

Mobile PhaseMeCN – 20%
BufferH2SO4 – 0.1%
Flow Rate1.0 ml/min
DetectionUV 250 nm
Class of CompoundsSalt
Analyzing CompoundsPotassium sorbate,Sorbic acid

HPLC Column Used

Newcrom BH, 4.6 x 150 mm, 5 µm, 100 A, dual ended