HPLC Method For Analysis Of  Patulin on Newcrom R1 Column

https://sielc.com/hplc-method-for-analysis-of-patulin
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Patulin.

Patulin is a toxic chemical contaminant produced by several species of mold, especially within Byssochlamys Aspergillus and Penicillium. It is the most common mycotoxin found in apples and apple-derived products such as juice, cider, compotes and other food intended for young children.

Patulin can be retained in HPLC on Newcrom R1 reverse-phase column with the simple isocratic mobile phase consisting of acetonitrile (MeCN, ACN), water and Phosphoric Acid (H3PO4) buffer. The analysis method can be UV detected at 275 nm.

Method Parameters

Mobile PhaseMeCN – 10%
BufferH3PO4 – 0.2%
Flow Rate0.5 ml/min
UV Detection275 nm
LOD0.002 ppm
Class of CompoundsHydrophilic
Analyzing CompoundsPatulin

HPLC Column Used

Newcrom R1, 3.2 x 100 mm, 5 µm, 100 A, dual ended