HPLC Method For Analysis Of Meldonium in Mildronate Tablet Dosage Form on Primesep 100 Column

https://sielc.com/hplc-method-for-analysis-of-meldonium
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Meldonium .

Meldonium is a medication with the chemical formula C 6 H 14 N 2 O 2 .  It is a popular treatment for ischemia in Eastern Europe.  Due to it’s metabolic modulating effects as well as it altering enzymatic reaction in the body it is considered to be a performance-enhancing drug, therefore, it has been banned by the World Anti-Doping Agency since 2016.  It works as a fatty acid oxydation inhibitor by inhibiting enzymes in the carnitine biosynthesis pathways.

Meldonium can be retained and analyzed using the Primesep 100 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a perchloric acid buffer. Detection is performed using UV.

Method Parameters

Mobile PhaseMeCN/H2O – 25/75%
BufferHClO4 – 0.2%
Flow Rate1.0 ml/min
DetectionUV, 205 nm
Class of CompoundsAmino acid, Drug
Analyzing CompoundsMeldonium

HPLC Column Used

Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended