HPLC Method for Analysis of Iprodione on Newcrom B Column on Alltesta™

https://sielc.com/hplc-method-for-analysis-of-iprodione
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Iprodione

Iprodione is a hydantoin fungicide and nematicide with the chemical formula C 13 H 13 Cl 2 N 3 O 3 . It is used on a wide variety of crops that are affected by fungal diseases such as botrytis bunch rot, brown rot, and Sclerotinia.  Approval of using it is different across the world.

Iprodione can be retained and analyzed using the Newcrom B stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with phosphoric acid as a buffer. Detection is performed using UV.

Method Parameters

Mobile PhaseMeCN – 70%
BufferSulfuric Acid
Flow Rate1.0 mL/min
DetectionUV 275 nm
Class of CompoundsDicarboximide fungicides
Analyzing CompoundsIprodione

HPLC Column Used

Newcrom B, 4.6 x 150 mm, 5 µm, 100 A, dual ended