· High Performance Liquid Chromatography (HPLC) Method for Analysis of Diglycolic Anhydride .
Diglycolic Anhydride is an compound with the chemical formula C 4 H 4 O 4 . It is used in synthesis of poly(ester-ether), poly(amide-ether), as well as as a starting material for production of bicyclic heterocyclic structures.
Diglycolic Anhydride can be retained and analyzed using the Primesep S2 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of acetonitrile (MeCN). Detection is performed using UV.
| Mobile Phase | MeCN- 100% |
|---|---|
| Buffer | No |
| Flow Rate | 1.0 ml/min |
| Detection | UV 210 nm |
| Class of Compounds | Anhydride |
| Analyzing Compounds | Diglycolic Anhydride |