· High Performance Liquid Chromatography (HPLC) Method for Analysis of Diglycolic Anhydride .

Diglycolic Anhydride is an compound with the chemical formula C 4 H 4 O 4 . It is used in synthesis of poly(ester-ether), poly(amide-ether), as well as as a starting material for production of bicyclic heterocyclic structures.

Diglycolic Anhydride can be retained and analyzed using the Primesep S2 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of acetonitrile (MeCN). Detection is performed using UV.

Method Parameters

Mobile PhaseMeCN- 100%
BufferNo
Flow Rate1.0 ml/min
DetectionUV 210 nm
Class of CompoundsAnhydride
Analyzing CompoundsDiglycolic Anhydride

HPLC Column Used

Primesep S2, 4.6×250 mm, 5 µm, 100A