HPLC Method for Analysis of Bis(2,2,6,6-tetramethyl-4-piperidyl) sebacate on Primesep B Column

https://sielc.com/hplc-method-for-analysis-of-bis2266-tetramethyl-4-piperidyl-sebacate-on-primesep-b-column
⬇ PDF Download

Chromatogram

Description

High Performance Liquid Chromatography (HPLC) Method for Analysis of Bis(2,2,6,6-tetramethyl-4-piperidyl) sebacate

Bis(2,2,6,6-tetramethyl-4-piperidyl) sebacate is an ultra violet hindered amine light stabilizer with the chemical formula C 28 H 52 N 2 O 4 . It is commonly used in plastics production and in the packaging and manufacturing of pharmaceutical products. It helps prevent oxidation caused by sunlight weathering in coatings and paint. In 2024, it was found added in illicitly sold fentanyl.

Bis(2,2,6,6-tetramethyl-4-piperidyl) sebacate can be retained and analyzed on a Primesep B reverse-phase column using a gradient analytical method with a mobile phase consisting of Acetonitrile (MeCN), water, and Sulfuric acid as the ionic modifier. This analysis method can be UV detected at 210 nm with high resolution and peak symmetry.

Method Parameters

Mobile PhaseGradient MeCN
BufferH2SO4 – 0.2%
Flow Rate1.0 ml/min
DetectionUV 210 nm
Class of CompoundsHydrophobic
Analyzing CompoundsBis(2,2,6,6-tetramethyl-4-piperidyl) sebacate

HPLC Column Used

Primesep B, 4.6 x 150 mm, 5 µm, 100 A, dual ended