HPLC Method for Analysis of Bicine on Primesep 100 Column

https://sielc.com/hplc-method-for-analysis-of-bicine
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Bicine .

Bicine is a common organic buffering agent wit the chemical formula C 6 H 13 NO 4 . It is one of many buffering agents known as “Good’s buffers,” and like many of them, Bicine is zwitterionic. It is used in low temperature conditions due to it’s stable pH and high solubility even at 0 o C.

Bicine can be retained and analyzed using the Primesep 100 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a sulfuric acid buffer. Detection is performed using UV.

Method Parameters

Mobile PhaseMeCN – 20%
BufferH2SO4 – 0.1%
Flow Rate1.0 ml/min
DetectionUV 200nm
Sample8 mg/ml in H2O
LOD*3 ppm
Class of CompoundsAmino acid
Analyzing CompoundsBicine

HPLC Column Used

Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended