HPLC Method For Analysis Of Alprazolam on Newcrom R1 Column

https://sielc.com/hplc-method-for-analysis-of-alprazolam
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Alprazolam

Alprazolam, also known as Xanax, is a perscription medication with the C 17 H 13 ClN 4 molecular formula. It is typically used to treat anxiety and panic disorders. It works through enhancing the effects of GABA, which is a calming neurotransmitter in the brain.

Alprazolam can be retained and analyzed using the Newcrom R1 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a phosphoric acid buffer. Detection is performed using UV.

Method Parameters

Mobile PhaseMeCN/H2O – 30/70%
BufferH3PO4 – 0.2%
Flow Rate0.5 ml/min
DetectionUV, 250 nm
Class of CompoundsDrug, Sedative, Hydrophobic, Neutral, Acid
Analyzing CompoundsAlprazolam

HPLC Column Used

Newcrom R1, 3.2 x 100 mm, 5 µm, 100 A, dual ended