HPLC Method for Analysis of Moroxydine on Primesep 100 Column

https://sielc.com/hplc-method-for-analysis-moroxydines
⬇ PDF Download

Chromatogram

Description

· HPLC Method for Analysis of Moroxydine on Primesep 100 Column by SIELC Technologies

Moroxydine is an antiviral medication with the chemical formula C 6 H 13 N 5 O . It is used in the treatment of influenza. It is believed to work through inhibiting the early stages of viral replication. You can find detailed UV spectra of Moroxydine and information about its various lambda maxima by visiting the following link.

Moroxydine can be retained and analyzed using a Primesep 100 mixed-mode stationary phase column. The analysis employs an isocratic method with a simple mobile phase comprising water, acetonitrile (MeCN), and ammonium formate as a buffer. This method allows for detection using UV at 240 nm

Method Parameters

Mobile PhaseMeCN/H2O 75/25%
BufferAmFm pH 3.0  – 30 mM
Flow Rate1.0 ml/min
DetectionUV 240 nm
Samples0.5 mg/ml
Injection volume2 µl
LOD*8 ppb
Class of CompoundsBiguanides
Analyzing CompoundsMoroxydine

HPLC Column Used

Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended