HPLC Method for Analysis of Bromophenol blue on Newcrom R1 Column

https://sielc.com/hplc-method-for-analysis-bromophenol-blue
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Bromophenol blue

Bromophenol blue (BPB) is a pH indicator and dye commonly used in laboratories, especially in gel electrophoresis and titrations.

pH Indicator Properties Color Changes: Below pH 3.0: Yellow Between pH 3.0 and 4.6: Greenish Above pH 4.6: Blue

Bromophenol blue can be retained and analyzed using the Newcrom R1 column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water, acetonitrile (MeCN), and sulfuric acid as a buffer. Detection is carried out using UV.

You can find detailed UV spectra of Bromophenol blue and information about its various lambda maxima by visiting the following link .

Method Parameters

Mobile Phase*Gradient MeCN – 30-70%, 10 min
BufferH2SO4 – 0.2%
Flow Rate0.5 ml/min
DetectionVIS 420, (λmax-200, 280, 424nm)
LOD**2 ppb
Class of CompoundsDye
Analyzing CompoundsBromophenol blue

HPLC Column Used

Newcrom R1, 3.2 x 100 mm, 5 µm, 100 A, dual ended