HPLC Method for Analysis of Isoprene on Newcrom R1 Column

https://sielc.com/hplc-elsd-method-for-analysis-of-isoprene
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Isoprene

Isoprene is a volatile organic compound with the molecular formula C 5 H 8 . It is most often found in plants and is released as protection from heat stress. It also helps with maintaining cell membrane stability. Isoprene emissions are also the reasons why Blue Ridge Mountains have a blue haze.

Isoprene can be retained and analyzed using the Newcrom R1 stationary phase column. The analysis utilizes a gradient method with a simple mobile phase consisting of water and acetonitrile (MeCN). Detection is performed using UV 222nm.

You can find detailed UV spectra of Isoprene and information about its various lambda maxima by visiting the following link.

Method Parameters

Mobile PhaseMeCN – 60%
BufferNone
Flow Rate1.0 ml/min
DetectionUV 222
LOD*35 ppb
Class of CompoundsTerpenoid
Analyzing CompoundsIsoprene

HPLC Column Used

Newcrom R1, 4.6 x 150 mm, 5 µm, 100 A, dual ended