HPLC Determination of Ractopamine on Primesep 100 Column

https://sielc.com/hplc-determination-of-ractopamine
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Ractopamine hydrochloride .

Ractopamine is a livestock feed additive designed to promote muscle synthesis and leanness. It is banned in many countries, excluding the US, Canada and Mexico. It can be retained on a Primesep 100 Column in about 7 minutes. Primesep 100 is a reverse-phase (RP) mixed-mode column with embedded acidic ion-pairing groups.

Method Parameters

Mobile PhaseMeCN/H2O – 20/80%
BufferH2SO4 – 0.3 %
Flow Rate1.0 ml/min
DetectionUV, 222 nm
Class of CompoundsDrug, Basic, Hydrophobic, Ionizable, Amine
Analyzing CompoundsRactopamine hydrochloride

HPLC Column Used