· High Performance Liquid Chromatography (HPLC) Method for Analysis of Prochloraz .

Prochloraz is an imidazole fungicide with the chemical formula C 15 H 16 Cl 3 N 3 O 2 . It is used widely across Europe, Asia, South America, and Australia, but is not registered for use in the United States. It works as an inhibitor of the enzyme lanosterol 14α-demethylase.

Prochloraz can be retained and analyzed using the Newcrom B stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with an ammonium formate buffer. Detection is performed using UV.

Method Parameters

Mobile PhaseMeCN/H2O – 50/50%
BufferAmFm pH 3.0 –  20 mM
Flow Rate1.0 ml/min
DetectionUV, 280 nm
Class of CompoundsHydrophobic, Herbicide, Pesticide
Analyzing CompoundsProchloraz

HPLC Column Used

Newcrom B, 4.6 x 150 mm, 5 µm, 100 A, dual ended