HPLC Determination of Imidazolidinyl urea on SHARC 1 Column

https://sielc.com/hplc-determination-of-imidazolidinyl-urea-on-sharc-1
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Imidazolidinyl urea .

Imidazolidinyl urea is a preservative with the chemical formula C 11 H 16 N 8 O 8 . used in cosmetics as due to its antimicrobial properties. It is active against a broad spectrum of gram positive and gram negative bacteria, yeast and molds.

Imidazolidinyl urea can be retained in HPLC using a SHARC 1 column, which uses hydrogen-bonding as a separation mechanism. The isocratic method uses a mobile phase of acetonitrile (ACN) and water. Imidazolidinyl urea can be monitored by low UV (210 nm), ELSD, CAD or LC/MS.

Method Parameters

Mobile PhaseMeCN/MeOH – 90/10%
BufferNo
Flow Rate1.0 ml/min
DetectionUV 210 nm
Class of CompoundsHydrophilic, Amide, Amine
Analyzing CompoundsImidazolidinyl urea

HPLC Column Used

SHARC 1, 4.6 x 150 mm, 5 µm, 100 A, dual ended