HPLC Determination of Gadolinium Acetate on Newcrom A Column

https://sielc.com/hplc-determination-of-gadolinium-acetate
⬇ PDF Download

Chromatogram

HPLC Determination of Gadolinium Acetate on Newcrom A Column_1230

HPLC Determination of Gadolinium Acetate on Newcrom A Column_1230

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Gadolinium acetate .

Gadolinium Acetate is an acetate salt with the chemical formula Gd(CH 3 COO) 3 .  It is a crystalline compound that is used as a key precursor for gadolinium-based materials, fabrication of MRI contrast agents, luminescent materials, perovskite solar cells, and thin-film coating.

Method Parameters

Mobile PhaseMeCN/H2O – 10/90%
BufferAmFm pH 3.0- 10 mM
Flow Rate1.0 ml/min
DetectionCAD (Corona) MS- compatible mobile phase
Class of CompoundsHydrophilic, Metal, Ion
Analyzing CompoundsGadolinium acetate

HPLC Column Used

Newcrom A, 4.6 x 150 mm, 5 µm, 100 A, dual ended