HPLC Determination of Fipronil on Newcrom R1 Column

https://sielc.com/hplc-determination-of-fipronil-on-newcrom-r1-column
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of Fipronil .

Fipronil is an insecticide with the chemical formula C 12 H 4 Cl 2 F 6 N 4 O S . It works through causing hyperexcitation of the nerves and muscles in contaminated insects.  In Texas, it is often used to control invasive species of ants and wasps. It is also an active ingredient in flea and tick treatments for pets. WHO classifies it as a Class II, moderately hazardous, pesticide.

Fipronil can be retained and analyzed using the Newcrom R1 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) without a buffer. Detection is performed using UV.

Method Parameters

Mobile PhaseMeCN/H2O
BufferNo
Flow Rate1.0 ml/min
DetectionUV 200 nm (MS-compatible mobile phase)
Class of CompoundsHydrophobic, Herbicide, Pesticide
Analyzing CompoundsFipronil

HPLC Column Used

Newcrom R1, 4.6 x 150 mm, 5 µm, 100 A, dual ended