HPLC Determination of EDTA on Newcrom BH Column

https://sielc.com/hplc-determination-of-edta-on-newcrom-bh-column
⬇ PDF Download

Chromatogram

Description

EDTA Standards Solution A: For the preparation of the EDTA standard solution, 5 mg of EDTA was accurately weighed and transferred into a 5 mL volumetric flask and dissolved in 0.001N NaOH water solution with sonication or magnetic stirrer mixing. Filtered The EDTA stock solution (1.0 mg/mL) should be stored in a cold dark place and can be used for a week to prepare standards of required concentration.

Iron(III) chloride Solution B: The standard stock solution of Iron(III) chloride (10 mg/ml) was prepared in water. 50 mg of FeCl3 was accurately weighed and transferred into a 5 mL volumetric flask and dissolved in water, with sonication if needed.

General procedure for Ferric EDTA complex analysis: To make a sample for analysis mix 100 µL Solution A (or unknown sample) with 100 µL Solution B and 800 µL of water.  Place this mixture in a plastic HPLC vial for analysis. Setup instrument and column according to the method provided.

Method Parameters

Mobile PhaseMeCN/H2O – 2/98%
BufferH2SO4 – 0.1%
Flow Rate1.0 ml/min
DetectionUV 260nm
Class of CompoundsAcid, Hydrophilic
Analyzing CompoundsEDTA

HPLC Column Used

Newcrom BH, 4.6×150 mm,  100A