· HPLC Method for Coenzyme q10 , Ubiquinol on Primesep D by SIELC Technologies · High Performance Liquid Chromatography (HPLC) Method for Analysis of Coenzyme q10 , Ubiquinol .
Coenzyme Q10 , also known as ubiquinone , is an enzyme found in animals and is used as a dietary supplement with the chemical formula C 59 H 90 O 4 . It acts as a redox carrier, which means that it transfers electrons in the mitochondria. The production of ubiquinone depletes with age. Ubiquinol is a reduced form of coenzyme Q10.
The two compounds are similar in structure and can be difficult to separate in reverse-phase HPLC. Using mixed-mode Primesep D column and the mobile phase with acetonitrile (ACN) and water, without the need for buffer, CoQ10 and ubiquinol can be isocratically separated and UV detected at 275nm.
| Mobile Phase | MeCN/H2O – 95/5% |
|---|---|
| Buffer | No |
| Flow Rate | 1.0 ml/min |
| Detection | UV 275 nm, Vis 435 nm |
| Class of Compounds | Hydrophobic |
| Analyzing Compounds | Coenzyme q10,Ubiquinol |