HPLC Determination of 1-pyridin-3-yl-ethylamine on Primesep 100 Column

https://sielc.com/hplc-determination-of-1-pyridin-3-yl-ethylamine-on-primesep-100-column
⬇ PDF Download

Chromatogram

Description

· High Performance Liquid Chromatography (HPLC) Method for Analysis of 1-pyridin-3-yl-ethylamine. · 1-pyridin-3-yl-ethylamine has the chemical formula C 7 H 10 N 2 .

1-pyridin-3-yl-ethylamine can be retained  on the Primesep 100 mixed-mode column using an isocratic analytical method with a simple mobile phase of water, acetonitrile (MeCN, ACN), and sulphuric acid (H2SO4) buffer. The analysis method can be UV detected at 260 nm.

Method Parameters

Mobile PhaseMeCN/H2O – 20/80%
BufferH2SO4 – 0.3%
Flow Rate1.0 ml/min
DetectionUV, 260 nm
Class of CompoundsAmine
Analyzing Compounds1-pyridin-3-yl-ethylamine

HPLC Column Used

Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended