USP Methods for the Analysis of Aminocaproic Acid using the Legacy L1 Column

https://sielc.com/Application-USP-Methods-for-the-Analysis-of-Aminocaproic-Acid-using-the-Legacy-L1-Column
⬇ PDF Download

Chromatogram

Description

Application Notes: Aminocaproic acid is a enzymatic inhibitor. According to USP methods, aminocaproic acid conatins not less than 98.5% and not more than 101.5% of aminocaproic acid calculated on a dried basis. The USP HPLC method for the separation of aminocaproic acid was developed on Legacy L1 column according to the US Pharmacopeia methodology. L1 classification is assigned to reversed-phase HPLC column containing C18 ligand. Support for the material is spherical silica gel with particles size 3-10 um and pore size of 100-120A. Resolution between critical pairs corresponds to rules and specifications of UPS.

Application Columns: Legacy L1 C18 HPLC column · Application compounds: Aminocaproic acid · Mobile phase: Sodium heptanesulfonate/H2O/MeOH /NaH2PO4 pH 2.2 · Detection technique: UV · Reference: USP35: NF30

Method Parameters

Mobile PhaseSodium heptanesulfonate/H2O/MeOH /NaH2PO4 pH 2.2
Flow Rate0.7 ml/min
DetectionUV, 210 nm
Class of CompoundsDrug, Clotting promoter, Hydrophilic, Ionizable
Analyzing CompoundsAminocaproic Acid

HPLC Column Used

Legacy L1, 4.6×150 mm, 5 µm, 100A