HPLC Separation of Antibiotics of Group in Mixed-Mode Chromatography

https://sielc.com/HPLC_Separation_of_Antibiotics_of_Group_in_Mixed-Mode_Chromatography
⬇ PDF Download

Chromatogram

Description

· Separation type: Liquid Chromatography Mixed-mode

Primesep D was used to separate the antibiotics marbofloxacin, norfloxacin, josamycin, and flumequine. Primesep D is a reverse-phase column with embedded basic ion-pairing groups designed for analysis of plasma. This m ethod is compatible with LC/MS and prep chromatography.

Method Parameters

Mobile PhaseGradient MeCN – 5-50%, 10 min, 3 min hold
BufferH2SO4 – 0.1%
Flow Rate1.0 ml/min
DetectionUV, 250 nm
Class of CompoundsAntibiotic, Drug, Hydrophobic, Ionizable
Analyzing CompoundsJosamycin, Norfloxacin, Marbofloxacin, Flumequine

HPLC Column Used

Primesep D, 4.6×150 mm, 5 µm, 100A