A unique approach for analysis of copper ions in a high salt content matrix was developed using EDTA as a visualization agent. Since copper ions can form colored complexes with EDTA, this property can be used for selective analysis of copper in the mixture with other cations, which do not form a colored complex with EDTA. Copper ion can be monitored by UV while other metals pass through the column undetected by UV. This method can be used to analyze copper ion at parts per million concentrations.

Method Parameters

Mobile PhaseMeCN/H2O – 30/70%
BufferAmAc, EDTA
Flow Rate1.0 ml/min
DetectionUV, 310 nm
Class of CompoundsIons
Analyzing CompoundsCopper

HPLC Column Used

Primesep SB, 4.6×50 mm, 5 µm, 100A