HPLC Separation of Organic Acids in HILIC and Anion-Exclusion Mode on Primesep S2 Column

https://sielc.com/Application-HPLC-Separation-of-Organic-Acids-in-HILIC-and-Anion-Exclusion-Mode-on-Primesep-S2-Column
⬇ PDF Download

Chromatogram

Description

Organic acids were separated on a HILIC/cation-exchange column in HILIC/anion-exclusion mode. This column can be used for analysis of polar compounds in HILIC mode. If compounds are ionizable, additional mode of interaction can be added (cation-exchange or anion-exclusion).

Method Parameters

Mobile PhaseMeCN/H2O – 85/15%
BufferAmAc pH 5.0 15 mM
Flow Rate1.0 ml/min
DetectionELSD, 50C UV 250 nm
Class of CompoundsNucleosides,  Hydrophilic, Ionizable
Analyzing CompoundsToluenesulfonic acid, Naphthalenedisulfonic acid, 3.5 DHBA, Ascorbic acid,

HPLC Column Used

Primesep S2, 4.6×150 mm, 5 µm, 100A