HPLC Analysis of Cheese Extract using Mixed-Mode Chromatography

https://sielc.com/Application-HPLC-Analysis-of-Cheese-Extract-Using-Mixed-Mode-Chromatography
⬇ PDF Download

Chromatogram

Description

Primesep 100 separates the components of a Swiss cheese extract by HPLC using cation exchange and reversed phase as retention mechanisms. The amino acids, glutamic acid and proline, as well as glycine betaine were resolved in less than 10 minutes. A mobile phase gradient of water, acetonitrile (MeCN, ACN,) and phosphoric acid (H3PO4) and ultraviolet (UV) detection was used.

Method Parameters

Mobile PhaseMeCN/H2O
BufferH3PO4
Flow Rate1.0 ml/min
DetectionUV 210 nm
Class of CompoundsDrug, Acid, Hydrophilic, Ionizable, Vitamin, Supplements
Analyzing CompoundsGlutamic acid,  Proline, Glycine betaine

HPLC Column Used

Primesep 100, 4.6×150 mm, 5 µm, 100A